Pressure Sensor Ifm PI2309


IFM CATEGORIES

Pressure Sensor Ifm pi2309

Pressure Sensor PI2309

Brand: Ifm
Product Code: PI-1-1BREZ02-MFRKG/US/ /P

  • Designed for high-temperature value usages in the food and beverage sector
  • Durable stainless steel enclosure featuring a hygienic flush-front design
  • Can be configured for either two-wire or three-wire connections
  • Simplified wiring configuration when functioning as a two-wire current loop
  • Built-in clamp diaphragm seal featuring a stainless steel membrane

Share on:

Product characteristics

Total count of inputs and outputs Count of digital outputs: 2; Count of analog outputs: 1
Measuring span
-1000...1000 mbar-14.5...14.5 psi-100...100 kPa
Connection for process Clamp DN50 (2") DIN 32676 (ISO 2852)
System gold-plated terminals
Application recessed installation compatible for the food and drink sector
Media viscous substances and fluids with suspended particulates; fluids and gases
Intermediate temperature value[°C] -20...200
Minimum force per unit area before bursting
30000 mbar435 psi3000 kPa
Pressure classification
10000 mbar145 psi1000 kPa
Kind of force per unit area relative force per unit area; vacuum
Damping process parameter dAP[s] 0...30
Damping for the analog output dAA[s] 0.01...99.99
Weight[g] 1200.3
Housing cylindrical
measurements[mm] Ø 50 / L = 163
Substance stainless steel (1.4404 / 316L); stainless steel (1.4435 / 316L); PBT; PEI; PFA
Substances (contacted components) stainless steel (1.4435 / 316L)
Minimum force per unit area cycles 10 million
Connection for process Clamp DN50 (2") DIN 32676 (ISO 2852)
Surface properties Ra/Rz of the wetted components < 0.38 / -
Transfer fluid for force per unit area transmission NEOBEE (food-safe, FDA approved)
diaphragm diameter[mm] 47.5

Interfaces

communication interface IO-Link
Mode of transmission COM2 (38,4 kBaud)
IO-Link version 1.0
SDCI specification IO-Link Communication Standard, Version 1.0, January 2009, Folder No: 10.002
Configurations without profile
SIO mode yes
Necessary master port category A
Analog data process 1
Data binary process 2
Minimum process cycle duration[ms] 2.3
Supported DeviceIDs>
Mode of operationDeviceID
default108
Surrounding temperature value[°C] -25...80
Comment on operating temperature value range
Intermediate temperature value: = 160 °C
Surrounding temperature value[°C] -25...65
Comment on operating temperature value range
Intermediate temperature value: = 200 °C
Temperature for storage[°C] -25...100
Safety feature IP 67; IP 68; IP 69K
Measuring span
-1000...1000 mbar-14.5...14.5 psi-100...100 kPa
Setpoint SP
-998...1000 mbar-14.45...14.5 psi-99.8...100 kPa
Reset threshold rP
-1000...998 mbar-14.5...14.45 psi-100...99.8 kPa
Analog initiation point
-1000...500 mbar-14.5...7.25 psi-100...50 kPa
Analog endpoint
-500...1000 mbar-7.25...14.5 psi-50...100 kPa
In increments of
1 mbar0.05 psi0.1 kPa
default configuration
SP1 = -500 mbarrP1 = -540 mbar
SP2 = 500 mbarrP2 = 460 mbar
ASP = -1000 mbarAEP = 1000 mbar

Inputs / outputs

Total count of inputs and outputs Count of digital outputs: 2; Count of analog outputs: 1
Minimum insulation resistance[MO] 100; (500 V DC)
Safety feature rating III
Reverse polarity safeguard yes
Built-in watchdog yes
Overall count of outputs 2
Output response output signal; analog output; IO-Link; (customizable)
Electrical engineering design PNP/NPN
Count of digital outputs 2
Output feature typically open / closed; (customizable)
Total analog outputs 1
Analog current output[mA] 4...20, invertible; (expandable)
Overcurrent protection yes
Method of overcurrent protection yes (momentary)
Overcurrent protection yes
Precision of switch points/span>[% of the span] < ± 0,2; (Turn down 1:1)
Reproducibility[% of the span] < ± 0,1; (with temperature value fluctuations < 10 K; Turn down 1:1)
Characteristics variation[% of the span] < ± 0,2; (Turn down 1:1 , linearity, incl. hysteresis and repeatability , limit value setting to DIN EN IEC 62828-1)
Linearity variation[% of the span] < ± 0,15; (Turn down 1:1)
Hysteresis variation[% of the span] < ± 0,15; (Turn down 1:1)
Extended stability[% of the span] < ± 0,1; (Turn down 1:1; per year)
Temperature coefficient value for the zero reference[% of the span / 10 K] < ± 0,3; (0...200 °C)
Temperature coefficient value for the range[% of the span / 10 K] < ± 0,15; (0...200 °C)

Download Specification for PI2309

Download PDF

RELATED PRODUCTS